Raman and Photoluminescence Spectroscopy

 

Contact:

Arnaud Magrez

 

Booking of the Raman Spectroscope

 

Renishaw inVia Reflex Raman Confocal microscope

  • Console Characteristics:

Research grade adapted Leica DM 2700 microscope with N-PLAN Objectives 5x (NA 0.25 – WD 17.6 mm), 50x (NA 0.75 – WD 0.37 mm), 50x (NA 0.50 – WD 8.20 mm), 100x (NA 0.85 – WD 0.33 mm), LWD PL-FLUOTARL100x (NA 0.75 – WD 4,7 mm).

 

  • Lasers:

All lasers are equipped with plasma lines rejection filter (PLRF) and kinematics laser mount. 

Wavelength Power Characteristics

405 nm

 

100 mW

 

Linear vertically polarized (>100:1). TEMoo M2<1.1. Line width <1,0 MHz, Spot size 0,7 mm +/- 0,05 (@ laser aperture) 

 

488 nm

 

100 mW

 

Linear vertically polarized (>100:1). TEMoo, M2<1.1. Line width <1,5 MHz, Spot size 0,65 mm – 0,75 mm (@ laser aperture)

 

532 nm

 

35 mW

 

Laser is vertically polarized (>100:1). TEMoo, M2<1.1. Line width <1 MHz, Spot size 0,7 mm (@ laser aperture) 

 

785 nm

 

125 mW

 

Linear vertically polarized (single mode). TEMoo. Line width 1,0 MHz, Spot size 0,7 mm (@ laser aperture)

 

 

  • Rayleigh filters:

Kinematically mounted, magnetically attached, Rayleigh line rejection filter (edge) set for the 405 nm excitation, allowing measurements to 250 cm-1. 

Kinematically mounted, magnetically attached, Rayleigh line rejection filter (edge) set for the 488, 532 and 785 nm excitation, allowing measurements to 50 cm-1. 

 

  • Monochromator and gratings characteristics:

Single monochromator system comprising the following kinematics gratings: 

Characteristics

Optimised for

300 l/mm

405, 488, 532, 785 nm

600 l/mm

405, 488, 532, 785 nm

1200 l/mm

785 nm

1800 l/mm 532 nm
2400 l/mm 405, 488 nm
3000 l/mm 405, 488, 532 nm

 

  • Point by point Imaging and XYZ stage: 

XYZ motorised microscope stage Renishaw HSES 100, with trackball, X and Y axis are encoded. (resolution 0.3 um, minimum step size 100 nm in X, Y and Z), for mapping experiments (point by point Imaging).